Gafchromic EBT2 film dosimetry in reflection mode with a novel plan-based calibration method

Published 17 Mar 2014 in physics.med-ph | (1403.4041v1)

Abstract: Purpose: A dosimetric system formed by Gafchromic EBT2 radiochromic film and Epson Expression 10000XL flatbed scanner was commissioned for dosimetry. In this paper, several open questions concerning the commissioning of radiochromic films for dosimetry were addressed: a) is it possible to employ this dosimetric system in reflection mode; b) if so, can the methods used in transmission mode also be used in reflection mode; c) is it possible to obtain accurate absolute dose measurements with Gafchromic EBT2 films; d) which calibration method should be followed; e) which calibration models should be used; f) does three-color channel dosimetry offer a significant improvement over single channel dosimetry. The purpose of this paper is to help clarify these questions. Conclusions: Reflection mode with Gafchromic EBT2 radiochromic film was found to be a viable alternative to transmission mode. The same methods that are used in transmission mode can be followed in reflection mode. A novel plan-based method was developed for calibration and multichannel dosimetry. This novel method offers increased robustness against film response inhomogeneities and reduces considerably the time required for calibration.

Summary

No one has generated a summary of this paper yet.

Paper to Video (Beta)

No one has generated a video about this paper yet.

Whiteboard

No one has generated a whiteboard explanation for this paper yet.

Open Problems

We haven't generated a list of open problems mentioned in this paper yet.

Continue Learning

We haven't generated follow-up questions for this paper yet.

Collections

Sign up for free to add this paper to one or more collections.