Papers
Topics
Authors
Recent
Search
2000 character limit reached

Tiny adiabatic-demagnetization refrigerator for a commercial superconducting quantum interference device magnetometer

Published 2 Dec 2016 in physics.ins-det, cond-mat.str-el, and cond-mat.supr-con | (1612.01376v1)

Abstract: A tiny adiabatic-demagnetization refrigerator (T-ADR) has been developed for a commercial superconducting quantum interference device magnetometer [Magnetic Property Measurement System (MPMS) from Quantum Design]. The whole T-ADR system is fit in a cylindrical space of the diameter $8.5$ mm and the length $250$ mm, and can be inserted into the narrow sample tube of MPMS. A sorption pump is self-contained in T-ADR, and hence no complex gas handling system is necessary. With the single crystalline Gd$3$Ga$_5$O${12}$ garnet ($\sim 2$ grams) used as a magnetic refrigerant, the routinely achievable lowest temperature is $\sim 0.56$ K. The lower detection limit for a magnetization anomaly is $\sim 1 \times 10{-7}$ emu, estimated from fluctuation of the measured magnetization. The background level is $\sim 5 \times 10{-5}$ emu below 2 K at $H = 100$ Oe, which is largely attributable to a contaminating paramagnetic signal from the magnetic refrigerant.

Citations (7)

Summary

No one has generated a summary of this paper yet.

Paper to Video (Beta)

No one has generated a video about this paper yet.

Whiteboard

No one has generated a whiteboard explanation for this paper yet.

Open Problems

We haven't generated a list of open problems mentioned in this paper yet.

Continue Learning

We haven't generated follow-up questions for this paper yet.

Collections

Sign up for free to add this paper to one or more collections.