Papers
Topics
Authors
Recent
Search
2000 character limit reached

High resolution ion trap time-of-flight mass spectrometer for cold trapped ion experiments

Published 21 Jul 2017 in physics.ins-det, physics.atom-ph, and physics.chem-ph | (1707.07036v1)

Abstract: Trapping molecular ions that have been sympathetically cooled with laser-cooled atomic ions is a useful platform for exploring cold ion chemistry. We designed and characterized a new experimental apparatus for probing chemical reaction dynamics between molecular cations and neutral radicals at temperatures below 1 K. The ions are trapped in a linear quadrupole radio-frequency trap and sympathetically cooled by co-trapped, laser-cooled, atomic ions. The ion trap is coupled to a time-of-flight mass spectrometer to readily identify product ion species, as well as to accurately determine trapped ion numbers. We discuss, and present in detail, the design of this ion trap time-of-flight mass spectrometer, as well as the electronics required for driving the trap and mass spectrometer. Furthermore, we measure the performance of this system, which yields mass resolutions of $m/\Delta{}m \geq 1100$ over a wide mass range, and discuss its relevance for future measurements in chemical reaction kinetics and dynamics.

Citations (31)

Summary

Whiteboard

No one has generated a whiteboard explanation for this paper yet.

Open Problems

We haven't generated a list of open problems mentioned in this paper yet.

Continue Learning

We haven't generated follow-up questions for this paper yet.

Collections

Sign up for free to add this paper to one or more collections.