Absolute ion detection efficiencies of microchannel plates and funnel microchannel plates for multi-coincidence detection

Published 22 Mar 2018 in physics.ins-det | (1803.08380v1)

Abstract: Modern momentum imaging techniques allow for the investigation of complex molecules in the gas phase by detection of several fragment ions in coincidence. For these studies, it is of great importance that the single-particle detection efficiency e is as high as possible, as the overall efficiency scales with e over n, i.e. the power of the number of detected particles. Here we present measured absolute detection efficiencies for protons of several micro-channel plates (MCPs), including efficiency enhanced "funnel MCPs". Furthermore, the relative detection efficiency for two-, three-, four-, and five-body fragmentation of CHBrClF has been examined. The "funnel" MCPs exhibit an efficiency of approx. 90 percent, gaining a factor of 24 (as compared to "normal" MCPs) in case of a five-fold ion coincidence detection.