Papers
Topics
Authors
Recent
Search
2000 character limit reached

Construction of a radiation beam scanner and investigation of volume averaging correction factor effects on beam-profile

Published 1 Nov 2022 in physics.ins-det and physics.med-ph | (2211.00413v1)

Abstract: In radiation beam-profile measurements, an accurate positioning of the detector with high position resolution is essential. For this purpose, we built a scanning device capable of moving a detector in three dimensions using mainly parts from a commercial 3D-printer. The accuracy and repeatability of movement was tested with caliper, laser displacement sensor, and repeated ${60}$Co beam-profile measurements in a water phantom. The results from the caliper and the sensor showed position accuracy for the scanner to be better than $\pm$150 $\mu$m. The standard deviation of the error in position from laser sensor measurements was approximately 30 $\mu$m and the beam profile scans showed a maximum deviation from the mean position of 50 $\mu$m. The effect of volume averaging correction factors on ${60}$Co beam-profile was investigated with two different sized ionization chambers. The differences in the profiles were reduced significantly after applying the correction factors.

Summary

No one has generated a summary of this paper yet.

Paper to Video (Beta)

No one has generated a video about this paper yet.

Whiteboard

No one has generated a whiteboard explanation for this paper yet.

Open Problems

We haven't generated a list of open problems mentioned in this paper yet.

Continue Learning

We haven't generated follow-up questions for this paper yet.

Collections

Sign up for free to add this paper to one or more collections.