A novel cryogenic VUV spectrofluorometer for the characterization of wavelength shifters
Abstract: We present a novel cryogenic VUV spectrofluorometer designed for the characterization of wavelength shifters (WLS) crucial for experiments based on liquid argon (LAr) scintillation light detection. Wavelength shifters like 1,1,4,4-tetraphenyl-1,3-butadiene (TPB) or polyethylene naphthalate (PEN) are used in these experiments to shift the VUV scintillation light to the visible region. Precise knowledge of the optical properties of the WLS at liquid argon's temperature (87 K) and LAr scintillation wavelength (128 nm) is necessary to model and understand the detector response. The cryogenic VUV spectrofluorometer was commissioned to measure the emission spectra and relative wavelength shifting efficiency (WLSE) of samples between 300 K and 87 K for VUV (128 nm to 190 nm) and UV (310 nm) excitation. New mitigation techniques for surface effects on cold WLS were established. As part of this work, the TPB-based wavelength shifting reflector (WLSR) featured in the neutrinoless double-beta decay experiment LEGEND-200 was characterized. The wavelength shifting efficiency was observed to increase by (54 +/- 5)% from room temperature (RT) to 87 K. PEN installed in LEGEND-200 was also characterized and a first measurement of the relative wavelength shifting efficiency and emission spectrum at RT and 87 K is presented. The WLSE of amorphous PEN was found to be enhanced by at least (37 +/- 4)% for excitation with 128 nm and by (52 +/- 3)% for UV excitation at 87 K compared to RT.
Paper Prompts
Sign up for free to create and run prompts on this paper using GPT-5.
Top Community Prompts
Collections
Sign up for free to add this paper to one or more collections.