Papers
Topics
Authors
Recent
Search
2000 character limit reached

A highly sensitive SF$_6$-based leak test system for JUNO 3-inch PMT underwater electronics boxes

Published 30 May 2025 in physics.ins-det and hep-ex | (2505.24142v1)

Abstract: A total of 25600 3-inch photomultiplier tubes (PMTs), along with their corresponding frontend electronics, have been installed at the Jiangmen Underground Neutrino Observatory (JUNO). These electronics are housed in 200 stainless steel boxes that operate underwater. To verify the sealing integrity of the underwater boxes following integration, we developed an SF$_6$-based leak test system, opting against the typical helium-based system due to helium's ability to penetrate the PMT glass. After a few hours of accumulating leaking SF$_6$ from the underwater boxes, a leak rate detection limit of $2.3\times{10}{-9}$~Pa$\cdot$m$3$/s in terms of SF$_6$ was achieved, corresponding to $1.65\times{10}{-8}$~Pa$\cdot$ m$3$/s helium equivalent. This meets the sensitivity requirement of 1$\times$10${-7}$~Pa$\cdot$m$3$/s. This system was critical in identifying and replacing a few cases of leaking underwater boxes before installation.

Summary

No one has generated a summary of this paper yet.

Paper to Video (Beta)

No one has generated a video about this paper yet.

Whiteboard

No one has generated a whiteboard explanation for this paper yet.

Open Problems

We found no open problems mentioned in this paper.

Continue Learning

We haven't generated follow-up questions for this paper yet.

Collections

Sign up for free to add this paper to one or more collections.

Tweets

Sign up for free to view the 2 tweets with 4 likes about this paper.